现在的位置: 首页时讯速递, 进展交流>正文
[NEJM发表论文]:一种血管外可植入复律除颤器的效果和安全性
2022年10月04日 时讯速递, 进展交流 [NEJM发表论文]:一种血管外可植入复律除颤器的效果和安全性已关闭评论

ORIGINAL ARTICLE

Efficacy and Safety of an Extravascular Implantable Cardioverter–Defibrillator

Paul Friedman, Francis Murgatroyd, Lucas V.A. Boersma, et al

N Engl J Med August 28, 2022
DOI: 10.1056/NEJMoa2206485

Abstract

BACKGROUND

The extravascular implantable cardioverter–defibrillator (ICD) has a single lead implanted substernally to enable pause-prevention pacing, antitachycardia pacing, and defibrillation energy similar to that of transvenous ICDs. The safety and efficacy of extravascular ICDs are not yet known.

METHODS

We conducted a prospective, single-group, nonrandomized, premarket global clinical study involving patients with a class I or IIa indication for an ICD, all of whom received an extravascular ICD system. The primary efficacy end point was successful defibrillation at implantation. The efficacy objective would be met if the lower boundary of the one-sided 97.5% confidence interval for the percentage of patients with successful defibrillation was greater than 88%. The primary safety end point was freedom from major system- or procedure-related complications at 6 months. The safety objective would be met if the lower boundary of the one-sided 97.5% confidence interval for the percentage of patients free from such complications was greater than 79%.

RESULTS

A total of 356 patients were enrolled, 316 of whom had an implantation attempt. Among the 302 patients in whom ventricular arrhythmia could be induced and who completed the defibrillation testing protocol, the percentage of patients with successful defibrillation was 98.7% (lower boundary of the one-sided 97.5% confidence interval [CI], 96.6%; P<0.001 for the comparison with the performance goal of 88%); 299 of 316 patients (94.6%) were discharged with a working ICD system. The Kaplan–Meier estimate of the percentage of patients free from major system- or procedure-related complications at 6 months was 92.6% (lower boundary of the one-sided 97.5% CI, 89.0%; P<0.001 for the comparison with the performance goal of 79%). No major intraprocedural complications were reported. At 6 months, 25 major complications were observed, in 23 of 316 patients (7.3%). The success rate of antitachycardia pacing, as assessed with generalized estimating equations, was 50.8% (95% CI, 23.3 to 77.8). A total of 29 patients received 118 inappropriate shocks for 81 arrhythmic episodes. Eight systems were explanted without extravascular ICD replacement over the 10.6-month mean follow-up period.

CONCLUSIONS

In this prospective global study, we found that extravascular ICDs were implanted safely and were able to detect and terminate induced ventricular arrhythmias at the time of implantation. (Funded by Medtronic; ClinicalTrials.gov number, NCT04060680. opens in new tab.)

抱歉!评论已关闭.

×
腾讯微博